ETC Request ID: STIS.sp.822390

Exposure time (seconds) = 253.0000 at wavelength 4751.00 Å

gives: SNR = 637.9707 (per resolution element)

gives: Time to Saturation (for a single exposure) = 365.23 seconds

Exposure time calculation HAD WARNINGS.

WARNING MESSAGE: "Electrons per pixel due to background (5) is less than the recommended threshold of 20 electrons to avoid poor charge transfer efficiency (CTE). We suggest you consider CTE mitigation strategies described in the STIS Instrument Handbook."


Table of Source and Noise Counts per Pixel


Plots


Detailed Information Count rate Total counts Associated noise
(e-/s) (e-) (e-)
Counts (box 7 pixels high) (1 pix x pix) (2.0 pix resel x pix)
    Source 806.496 408,086.90 638.82
    Background 0.142 71.62 8.46
        Sky 0.009 4.32 2.08
        Dark Current 0.133 67.30 8.20
    Read Out 31.80
Total in selected region 806.637 408,158.52 639.66
Brightest Pixel (single exposure) (at 4571.30 Å) 328.557 83,125.01

Global Count Rates
    Source only 555,848.991 140,629,794.739
    Background (sky+dark) 20,731.853 5,245,158.75

Encircled energy fraction 0.93


Instrument name: STIS
Mode: spectroscopic
Detector: CCD
Central Wavelength:
Grating: [G430L] R ~ 500
Aperture: 52X2
Gain: 4.0
# Frames: 1
Binning: dispersion: 1, spatial: 1
Dark:


Target: [point source]
Spectrum: Kurucz Models F8V 6200 4.4
Extinction E(B-V): None
Normalization: Renormalized to vegamag = 10.3 in filter Johnson/V
Redshift: None
Emission Lines: None


Selected background:
  Sky Background:
    Level:

Table of Source and Noise Counts per Pixel


Plots

These results were computed using: ETC version 24.2