ETC Request ID: STIS.sp.636595

Exposure time (seconds) = 368.0000 at wavelength 2376.00 Å

gives: SNR = 142.9267 (per resolution element)

gives: Time to Saturation (for a single exposure) = 1,703.42 seconds

Exposure time calculation HAD WARNINGS.

WARNING MESSAGE: "Electrons per pixel due to background (3.1) is less than the recommended threshold of 20 electrons to avoid poor charge transfer efficiency (CTE). We suggest you consider CTE mitigation strategies described in the STIS Instrument Handbook."


Table of Source and Noise Counts per Pixel


Plots


Detailed Information Count rate Total counts Associated noise
(e-/s) (e-) (e-)
Counts (box 7 pixels high) (1 pix x pix) (2.0 pix resel x pix)
    Source 29.011 21,351.92 146.12
    Background 0.119 87.59 9.36
        Sky 9.787e-06 7.20e-03 0.08
        Dark Current 0.119 87.58 9.36
    Read Out 29.63
Total in selected region 29.130 21,439.51 149.39
Brightest Pixel (single exposure) (at 3052.10 Å) 19.373 3,564.60

Global Count Rates
    Source only 33,990.190 12,508,389.800
    Background (sky+dark) 17,856.139 6,571,059.26

Encircled energy fraction 0.95


Instrument name: STIS
Mode: spectroscopic
Detector: CCD
Central Wavelength:
Grating: [G230LB] R ~ 700
Aperture: 52X2
Gain: 1.0
# Frames: 2
Binning: dispersion: 1, spatial: 1
Dark:


Target: [point source]
Spectrum: Standard HST Star Spectra Grw+70°5824 (DA3, 12.77) [1140.56, 24976.7 Å]
Extinction E(B-V): None
Normalization: None
Redshift: None
Emission Lines: None


Selected background:
  Sky Background:
    Level:

Table of Source and Noise Counts per Pixel


Plots

These results were computed using: ETC version 22.2