ETC Request ID: STIS.sp.1365947

Exposure time (seconds) = 2,580.0000 at wavelength 2400.00 Å

gives: SNR = 7.1767 (per resolution element)


Table of Source and Noise Counts per Pixel


Plots


Detailed Information Count rate Total counts Associated noise
(counts/s) (counts) (counts)
Counts (box 7 pixels high) (1 pix x 7 pix) (2 pix resel x 7 pix)
    Source 0.018 91.96 9.59
    Background 0.014 72.24 8.50
        Sky 5.370e-07 2.77e-03 0.05
        Dark Current 0.014 72.24 8.50
Total in selected region 0.032 164.21 12.81
Brightest Pixel (single exposure) (at 2465.60 Å) 0.009 23.93
Count rate entire detector 2,112.154

Buffer Time (sec) 946.90
For APT purposes, the recommended buffer time should be 4/5 of the buffer time calculated above. Please refer to the STIS Instrument Handbook for more details.

Global Count Rates
    Source only 13.284 34,273.796
    Background (sky+dark) 2,098.870 5,415,084.61

Encircled energy fraction 0.86


Instrument name: STIS
Mode: spectroscopic
Detector: NUVMAMA
Central Wavelength:
Grating: [G230L] R ~ 500
Aperture: 52X0.2


Target: [point source]
Spectrum: Black body at 20000.0K
Extinction E(B-V): None
Normalization: Renormalized to abmag = 20.4 in filter WFC3/UVIS1/F200LP
Redshift: z=0.088
Emission Lines: None


Selected background:
  Sky Background:
    Level:

Table of Source and Noise Counts per Pixel


Plots

These results were computed using: ETC version 27.2