ETC Request ID: STIS.sp.1168819

Exposure time (seconds) = 3,000.0000 at wavelength 2300.00 Å

gives: SNR = 188.7593 (per resolution element)


Table of Source and Noise Counts per Pixel


Plots


Detailed Information Count rate Total counts Associated noise
(counts/s) (counts) (counts)
Counts (box 7 pixels high) (1 pix x 7 pix) (2 pix resel x 7 pix)
    Source 5.949 35,692.98 188.93
    Background 0.011 63.00 7.94
        Sky 6.083e-07 3.65e-03 0.06
        Dark Current 0.011 63.00 7.94
Total in selected region 5.959 35,755.99 189.09
Brightest Pixel (single exposure) (at 2606.40 Å) 2.897 8,691.93
Count rate entire detector 6,712.205

Buffer Time (sec) 297.96
For APT purposes, the recommended buffer time should be 4/5 of the buffer time calculated above. Please refer to the STIS Instrument Handbook for more details.

Global Count Rates
    Source only 5,137.620 15,412,861.178
    Background (sky+dark) 1,574.585 4,723,754.99

Encircled energy fraction 0.85


Instrument name: STIS
Mode: spectroscopic
Detector: NUVMAMA
Central Wavelength:
Grating: [G230L] R ~ 500
Aperture: 52X0.2


Target: [point source]
Spectrum: Flat in units of Fλ
Extinction E(B-V): None
Normalization: Flux 6e-14 flam at 1900.0 Å
Redshift: None
Emission Lines: None


Selected background:
  Sky Background:
    Level:

Table of Source and Noise Counts per Pixel


Plots

These results were computed using: ETC version 26.1