ETC Request ID: STIS.sp.1035935

Requested Signal/Noise Ratio = 500.000 at wavelength 9500.00 Å (per resolution element)

gives: Time = 50.4751 seconds

gives: Time to Saturation (for a single exposure) = 69.91 seconds

Exposure time calculation HAD WARNINGS.

WARNING MESSAGE: "Electrons per pixel due to background (0.45) is less than the recommended threshold of 20 electrons to avoid poor charge transfer efficiency (CTE). We suggest you consider CTE mitigation strategies described in the STIS Instrument Handbook."


Table of Source and Noise Counts per Pixel


Plots


Detailed Information Count rate Total counts Associated noise
(e-/s) (e-) (e-)
Counts (box 7 pixels high) (1 pix x 7 pix) (2 pix resel x 7 pix)
    Source 2,495.106 251,881.29 501.88
    Background 0.126 12.72 3.57
        Sky 4.701e-05 4.75e-03 0.07
        Dark Current 0.126 12.72 3.57
    Read Out 43.39
Total in selected region 2,495.232 251,894.01 503.76
Brightest Pixel (single exposure) (at 9050.55 Å) 1,716.533 43,321.05

Global Count Rates
    Source only 4,487,174.554 226,490,456.023
    Background (sky+dark) 18,882.328 953,086.88

Encircled energy fraction 0.68


Instrument name: STIS
Mode: spectroscopic
Detector: CCD
Central Wavelength: 9336
Grating: [G750M] R ~ 5000
Aperture: 52X0.1
Gain: 4.0
# Frames: 2
Binning: dispersion: 1, spatial: 1
Dark: medium


Target: [point source]
Spectrum: Bruzual Synthetic Stellar Spectra O9 III
Extinction E(B-V): Milky Way Diffuse (Rv=3.1) = 0.1 applied before normalization
Normalization: Renormalized to vegamag = 4.7 in filter Johnson/J
Redshift: None
Emission Lines: None


Selected background:
  Sky Background:
    Level:

Table of Source and Noise Counts per Pixel


Plots

These results were computed using: ETC version 25.2