ETC Request ID: STIS.sp.1003499

Exposure time (seconds) = 1,750.0000 at wavelength 1900.00 Å

gives: SNR = 10.4389 (per resolution element)


Table of Source and Noise Counts per Pixel


Plots


Detailed Information Count rate Total counts Associated noise
(counts/s) (counts) (counts)
Counts (box 7 pixels high) (1 pix x 7 pix) (2 pix resel x 7 pix)
    Source 0.039 137.99 11.75
    Background 0.011 36.75 6.06
        Sky 2.664e-08 9.32e-05 9.66e-03
        Dark Current 0.011 36.75 6.06
Total in selected region 0.050 174.74 13.22
Brightest Pixel (single exposure) (at 2398.40 Å) 0.027 47.17
Count rate entire detector 1,623.573

Buffer Time (sec) 1,231.85
For APT purposes, the recommended buffer time should be 4/5 of the buffer time calculated above. Please refer to the STIS Instrument Handbook for more details.

Global Count Rates
    Source only 48.967 85,691.993
    Background (sky+dark) 1,574.606 2,755,561.24

Encircled energy fraction 0.84


Instrument name: STIS
Mode: spectroscopic
Detector: NUVMAMA
Central Wavelength:
Grating: [G230L] R ~ 500
Aperture: 52X0.2


Target: [point source]
Spectrum: Black body at 30000.0K
Extinction E(B-V): None
Normalization: Renormalized to abmag = 20.0 in filter Sloan/R
Redshift: None
Emission Lines: None


Selected background:
  Sky Background:
    Level:

Table of Source and Noise Counts per Pixel


Plots

These results were computed using: ETC version 25.2