ETC Request ID: STIS.sp.826937

Exposure time (seconds) = 475.0000 at wavelength 2800.00 Å

gives: SNR = 42.2754 (per resolution element)

Exposure time calculation HAD WARNINGS.

WARNING MESSAGE: MAMA Bright limits: total count rate per image 95999.9 exceeds limit for irregularly-variable sources 40 percent of the bright limit 200000 counts per second

WARNING MESSAGE: Buffer time 20.8333 is less than minimum 99 seconds.


Table of Source and Noise Counts per Pixel


Plots


Detailed Information Count rate Total counts Associated noise
(counts/s) (counts) (counts)
Counts (box 7 pixels high) (1 pix x pix) (2.0 pix resel x pix)
    Source 1.942 1,844.71 42.95
    Background 0.011 9.98 3.16
        Sky 9.935e-11 9.44e-08 3.07e-04
        Dark Current 0.011 9.98 3.16
Total in selected region 1.952 1,854.69 43.64
Brightest Pixel (single exposure) (at 2827.11 Å) 1.663 789.83
Count rate entire detector 95,999.924

Buffer Time (sec) 20.83
For APT purposes, the recommended buffer time should be 4/5 of the buffer time calculated above. Please refer to the STIS Instrument Handbook for more details.

Global Count Rates
    Source only 83,786.225 39,798,456.838
    Background (sky+dark) 1,572.865 747,110.94

Encircled energy fraction 0.93


Instrument name: STIS
Mode: spectroscopic
Detector: NUVMAMA
Central Wavelength: 2812
Grating: [E230H] R ~ 114000
Aperture: 31X0.05NDA


Target: [point source]
Spectrum: Castelli-Kurucz Models A5V 8250 4.0
Extinction E(B-V): None
Normalization: Renormalized to vegamag = 3.86 in filter Johnson/V
Redshift: None
Emission Lines: None


Selected background:
  Sky Background:
    Level:

Table of Source and Noise Counts per Pixel


Plots

These results were computed using: ETC version 24.2