ETC Request ID: STIS.sp.826455

Requested Signal/Noise Ratio = 10.000 at wavelength 1718.00 Å (per resolution element)

gives: Time = 547.4664 seconds


Table of Source and Noise Counts per Pixel


Plots


Detailed Information Count rate Total counts Associated noise
(counts/s) (counts) (counts)
Counts (box 7 pixels high) (1 pix x pix) (2.0 pix resel x pix)
    Source 0.114 125.11 11.19
    Background 0.011 11.50 3.39
        Sky 2.641e-11 2.89e-08 1.70e-04
        Dark Current 0.011 11.50 3.39
Total in selected region 0.125 136.60 12.51
Brightest Pixel (single exposure) (at 2319.51 Å) 0.343 187.55
Count rate entire detector 12,060.244

Buffer Time (sec) 165.83
For APT purposes, the recommended buffer time should be 4/5 of the buffer time calculated above. Please refer to the STIS Instrument Handbook for more details.

Global Count Rates
    Source only 8,257.784 4,520,859.118
    Background (sky+dark) 1,572.864 861,090.22

Encircled energy fraction 0.92


Instrument name: STIS
Mode: spectroscopic
Detector: NUVMAMA
Central Wavelength: 1978
Grating: [E230M] R ~ 30000
Aperture: 0.2X0.2


Target: [point source]
Spectrum: Castelli-Kurucz Models O7V 37000 4.0
Extinction E(B-V): LMC 30DorShell (Rv=2.76) = 0.56 applied before normalization
Normalization: Renormalized to vegamag = 9.67 in filter Johnson/V
Redshift: None
Emission Lines: None


Selected background:
  Sky Background:
    Level:

Table of Source and Noise Counts per Pixel


Plots

These results were computed using: ETC version 24.2