ETC Request ID: STIS.im.825280

Exposure time (seconds) = 0.5000

gives: SNR = 457.8925

gives: Time to Saturation (for a single exposure) = 1.28 seconds

gives: Optimum SNR = 517.5776

Exposure time calculation HAD WARNINGS.

WARNING MESSAGE: "Electrons per pixel due to background (0.023) is less than the recommended threshold of 20 electrons to avoid poor charge transfer efficiency (CTE). We suggest you consider CTE mitigation strategies described in the STIS Instrument Handbook."



Plots


Detailed Information Count rate Total counts Associated noise
(e-/s) (e-) (e-)
Counts (circle with radius 0.265 arcsec)
    Source 431,825.299 215,912.65 464.66
    Background 5.987 2.99 1.73
        Sky 4.296 2.15 1.47
        Dark Current 1.691 0.85 0.92
    Read Out 80.19
Total in selected region 431,831.286 215,915.64 471.54
Brightest Pixel (single exposure) 93,999.205 46,999.60

Global Count Rates
    Source only 539,781.624 269,890.812
    Background (sky+dark) 47,758.524 23,879.26

Background per unit area
    Sky Background (arcsec-2) 19.417 9.708
    Dark Current (arcsec-2) 7.600 3.800

Extraction region
circle with radius 0.265 arcsec
    Effective Sky Area (pixels) 88.49
    Fraction of flux 0.80
    Fraction of flux in brightest pixel 0.17
Effective Wavelength 7604.25 Å


Instrument name: STIS
Mode: imaging
Detector: CCD
Aperture: F28X50LP
Gain: 4.0
# Frames: 1
Binning: 1
Dark:


Target: [point source]
Spectrum: Pickles Models M0V 3801.89K
Extinction E(B-V): None
Normalization: Renormalized to vegamag = 11.88 in filter Johnson/V
Redshift: None
Emission Lines: None


Selected background:
  Sky Background:
    Level:


Plots

These results were computed using: ETC version 24.2