ETC Request ID: STIS.sp.823154

Exposure time (seconds) = 88.0000 at wavelength 5828.00 Å

gives: SNR = 979.5559 (per resolution element)

gives: Time to Saturation (for a single exposure) = 45.33 seconds

Exposure time calculation HAD WARNINGS.

WARNING MESSAGE: CCD Full Well limits: total counts per native pixel 232950 exceeds bright limit 120000 for selected gain

WARNING MESSAGE: "Electrons per pixel due to background (1.8) is less than the recommended threshold of 20 electrons to avoid poor charge transfer efficiency (CTE). We suggest you consider CTE mitigation strategies described in the STIS Instrument Handbook."


Table of Source and Noise Counts per Pixel


Plots


Detailed Information Count rate Total counts Associated noise
(e-/s) (e-) (e-)
Counts (box 7 pixels high) (1 pix x pix) (2.0 pix resel x pix)
    Source 5,457.761 960,565.89 980.08
    Background 0.147 25.83 5.08
        Sky 0.014 2.42 1.56
        Dark Current 0.133 23.41 4.84
    Read Out 31.80
Total in selected region 5,457.908 960,591.72 980.61
Brightest Pixel (single exposure) (at 5828.90 Å) 2,647.154 232,949.51

Global Count Rates
    Source only 3,454,031.726 303,954,791.878
    Background (sky+dark) 21,349.248 1,878,733.86

Encircled energy fraction 0.93


Instrument name: STIS
Mode: spectroscopic
Detector: CCD
Central Wavelength:
Grating: [G750L] R ~ 500
Aperture: 52X2
Gain: 4.0
# Frames: 1
Binning: dispersion: 1, spatial: 1
Dark:


Target: [point source]
Spectrum: Kurucz Models F2V 6890 4.3
Extinction E(B-V): None
Normalization: Renormalized to vegamag = 8.54 in filter Johnson/V
Redshift: None
Emission Lines: None


Selected background:
  Sky Background:
    Level:

Table of Source and Noise Counts per Pixel


Plots

These results were computed using: ETC version 24.2