All of the supported slits for this Cycle are available on the ETCs. The pull- down menu gives the supported slits and filters for the chosen grating/prism. Since the filters and the slits are in the same wheel, you can choose either a slit or a filter. In case of the neutral density filters, the pull-down menu gives the density of the filter in “log10” units, i.e. ND=1 corresponds to an attenuation factor of 10, ND=2 corresponds to an attenuation factor of 100, etc. In the case of the slits, the pull-down menu gives the slit height and slit width in arcsec. To choose the best slit width, you may need to know the number of detector pixels corresponding to the slit-width. The table below gives the plate scales for different gratings.
Grating |
Plate_Scale_(arcsec/pixel) |
|
---|---|---|
Along_Dispersion |
Along_Slit |
|
First Order CCD |
0.05 |
0.05 |
G140L, G230L |
0.0244 |
0.0244 |
G140M, G230M |
0.0290 |
0.0290 |
E230M |
0.0290 |
0.0290 |
E140M |
0.0290 |
0.0290 |
E230H |
0.0290 |
0.0290 |
E140H |
0.0290 |
0.0290 |
The ETC uses the current MAMA countrate restrictions, which are taken from the STIS Instrument Handbook. If the observation will exceed any of the countrate restrictions, a warning message will be given in the output page. This is particularly useful to check if the observations will exceed the bright-object protection (BOP) limit of the MAMA detector. The countrate restrictions for the detectors are as follows:
Gain |
Saturation Level |
Full Well |
||
---|---|---|---|---|
(electrons) |
(ADU) |
(electrons) |
(ADU) |
|
1 |
33,000 |
30,000 |
120,000 [1] |
120,000 [1] |
4 |
120,000 [1] |
30,000 |
120,000 [1] |
30,000 |
Detector |
Target |
Local Limit |
Global Limit |
||
---|---|---|---|---|---|
(First-order + Prism Modes) |
(Echelle Modes) |
(Imaging Mode) |
|||
(counts/sec/pixel) |
(counts/sec) |
(counts/sec) |
(counts/sec) |
||
NUV-MAMA |
Predictable |
100 |
30,000 |
200,000 |
200,000 |
FUV-MAMA |
Predictable |
75 |
30,000 |
200,000 |
|
NUV-MAMA |
Irregularly Variable |
75 |
12,000 |
80,000 |
|
FUV-MAMA |
Irregularly Variable |
75 |
12,000 |
80,000 |
In the case of the echelle modes, there is some extra noise because of the scattered light which runs across the orders. The updated version of the ETC takes this extra noise into account (to a first approximation) in the calculation of the S/N ratio. The global countrate estimates take the scattering into account.
For imaging mode the peak countrates mentioned in the output need some explanation. The peak countrates are used only to check for the MAMA bright object protection issues. Since the health of the detector sometimes relies on our ability to predict the peak countrate in a given observation, we have been a bit conservative in our estimates of peak countrates, particularly for the MAMAs. At present, the calculation of peak count rates for the point sources assume that the encircled energy in the central pixel is 30% in case of the CCD, and 25% in case of the MAMAs. This can be sometimes over conservative and can be slightly larger than a factor of 2 for some MAMA modes. A more accurate algorithm will be added in the future, which will take the spectral shape of a given source and the filter combination into account to calculate the appropriate percentage of enclosed energy in the central pixel.
The CCD has 3 binning factors: x1, x2 and x4. In Imaging mode the binning factors must be the same in each detector dimension, but for spectroscopy the factors can be different. Since the readnoise of the CCD applies only to a ‘binned’ pixel, using a binning factor greater than one can reduce the overall noise in some cases. However, this comes at the price of degraded spatial (or spectral) resolution. The readnoise and saturation characteristics of the CCD are given in the Table below. (Note that the MAMA detectors have no readnoise, so the binning option is ignored in these cases.)
STIS ACQ and ACQ/PEAK observations are always done using an unbinned subarray and a CCD gain setting of 4. For these parameters, the expected dark current (for low, medium (default), and high values) and read noise are similar to the CCD (Gain=4) line in the table below.
Detector |
Dark |
Read_Noise |
---|---|---|
CCD (Gain=1) |
[0.027, 0.033, 0.039] electrons/pixel/s [2] |
6.5 e- |
CCD (Gain=4) |
[0.027, 0.033, 0.039] electrons/pixel/s |
8.9 e- |
NUV-MAMA |
1.1 x 10-3 counts/pixel/s |
|
FUV-MAMA |
3.0 x 10-5 counts/pixel/s (no glow component) Typical for the first 2 orbits since FUV detector turn-on. |
|
|
7.5 x 10-5 counts/pixel/s (low glow component) [3] Typical for 2-3 orbits in the glow region or 4-5 orbits at the D1 pseudoaperture position. |
|
|
1.5 x 10-4 counts/pixel/s (medium glow component) Typical for 4 orbits in the glow region or 5+ orbits at the D1 pseudoaperture position. |
|
|
4.0 x 10-4 counts/pixel/s (high glow component) Typical for 5+ orbits in the glow region. |
|
For the top (E1 position), middle, and bottom of the CCD detector, respectively.
Observations cannot be guaranteed to be scheduled early in the detector’s high voltage time on. However, much of the time-variable glow component of the dark rate may be avoided by observing point sources at the D1 pseudoaperture position. See STIS IHB 4.2.3 for more details.
Setting the CCD gain to 4 is useful if you are expecting a large number of counts. For faint sources, where the counts are expected to be lower, gain=1 may be preferable since the readnoise (in counts) in that case is lower.
To avoid an excessive number of cosmic ray detections with the CCD, longer exposures should be split (CR-split) in order to:
keep the number of detected cosmic rays low
be able to remove the cosmic rays during data reduction
Using CR-split increases the effective readnoise, which is taken into account in the exposure time calculations. The default CR-split is 2.
For extended source acquisitions, the user sets CHECKBOX=n, where n must be an odd number between 3 and 105: the checkbox will then have dimension n x n pixels. CHECKBOX should be set to the minimum size which ensures that the brightest checkbox will be the one centered on the region of interest (i.e., if your object is peaked within a region of 1 arcsecond, set CHECKBOX=21 [= (1 arcsecond) / (0.05 arcsecond pixel-1) + 1]. The maximum checkbox is 105 pixels on a side, or ~5X5 arcseconds. The subarray used for a diffuse-source acquisition target image is CHECKBOX+101 pixels on a side. The STIS Target Acquisition Simulator can be used to determine the optimal CHECKBOX size.